Login / Signup

Robotic inguinal hernia repair with the new Hugo RAS TM system: first worldwide case series report.

Yoav MintzAlon J PikarskyRonit BrodieRam ElazariBrigitte HelouGad Marom
Published in: Minimally invasive therapy & allied technologies : MITAT : official journal of the Society for Minimally Invasive Therapy (2023)
Mean docking time was 9.5 min and mean console time was 50.3 min and 74.7 min for unilateral and bilateral inguinal hernia repair, respectively. No intraoperative or postoperative complications occurred. There was one conversion to laparoscopic surgery due to a technical issue with the robot. Conclusions: The modular design of independent arms seems to enhance maneuverability of the instruments and contribute to fewer arm collisions. Further clinical data and experience with this new surgical system are necessary to understand if this new design has advantages over the standard robotic platforms.
Keyphrases