Login / Signup

An efficient optimization method to improve the measuring accuracy of oxygen saturation by using triangular wave optical signal.

Gang LiYue YuCui ZhangLing Lin
Published in: The Review of scientific instruments (2018)
The oxygen saturation is one of the important parameters to evaluate human health. This paper presents an efficient optimization method that can improve the accuracy of oxygen saturation measurement, which employs an optical frequency division triangular wave signal as the excitation signal to obtain dynamic spectrum and calculate oxygen saturation. In comparison to the traditional method measured RMSE (root mean square error) of SpO2 which is 0.1705, this proposed method significantly reduced the measured RMSE which is 0.0965. It is notable that the accuracy of oxygen saturation measurement has been improved significantly. The method can simplify the circuit and bring down the demand of elements. Furthermore, it has a great reference value on improving the signal to noise ratio of other physiological signals.
Keyphrases
  • human health
  • risk assessment
  • high resolution
  • climate change
  • high speed
  • air pollution
  • mass spectrometry