Laboratory Evaluation of Linearity, Repeatability, and Hematocrit Interference With an Internet-Enabled Blood Glucose Meter.
Filiz DemircikValeria KirschSanja RamljakMario VoggAnke H PfütznerAndreas PfütznerPublished in: Journal of diabetes science and technology (2019)
The BGM met repeatability requirements but demonstrated a significant measurement bias in the low BG range. In addition, it failed the ISO15197:2015 criteria for hematocrit interference.