Login / Signup

Finite Element Assessment of the Screw and Cement Technique in Total Knee Arthroplasty.

Chong ZhengHai-Yang MaYin-Qiao DuJing-Yang SunJi-Wei LuoDong-Bin QuYong-Gang Zhou
Published in: BioMed research international (2020)
(1) The screw and cement technique is superior to the bone cement-only technique. For tibial defects in which the defect area comprises a large percentage but the depth is less than 5 mm, the screw and cement technique is recommended. (2) Vertical screws can achieve better stability than oblique screws. (3) Screws should be used in moderation for different defects; more is not always better.
Keyphrases
  • total knee arthroplasty
  • finite element
  • total hip
  • finite element analysis
  • soft tissue
  • bone loss
  • anterior cruciate ligament reconstruction