CT acquisition parameter selection in the real world: impacts on radiation dose and variation amongst 155 institutions.
Yifei WangPhilip ChuTimothy P SzczykutowiczCarly StewartRebecca Smith-BindmanPublished in: European radiology (2023)
• The relationship between CT acquisition parameter selection and radiation dose roughly coincided with established theoretical understanding. • CT acquisition parameters differ from each other in frequency and magnitude of manipulation, with beam energy (kV) being rarely manipulated. • Beam energy (kV) has the potential to substantially impact radiation dose, but because it is rarely manipulated, it is the least impactful CT acquisition parameter affecting radiation dose in practice.