Effect of MRI acquisition parameters on accuracy and precision of phase-contrast measurements in a small-lumen vessel phantom.
Maria Correia de VerdierJohan BerglundJohan WikströmPublished in: European radiology experimental (2024)
• PC-MRI measures flow and velocity in a small lumen phantom with high precision but overestimates flow rate. • The precision of PC-MRI measurements matches the precision of standard reference for flow rate measurements. • Optimizing PC-MRI settings can enhance accuracy and precision in flow rate and velocity measurements.