Accuracy of enteral syringes with commonly prescribed paediatric liquid medicines.
Sara Arenas-LópezKaruna GurungShane M TibbyMiguel Ángel Calleja HernándezCatherine TuleuPublished in: Archives of disease in childhood (2017)
Dosing was inaccurate when small volumes were used across all syringe sizes and brands. These reflect volumes used in clinical practice. Administration error could potentially be reduced by (1) clinicians using syringes appropriate to dosing volumes and (2) manufacturers revising formulation concentrations for drugs.