Diagnostic performance of CT versus MRI Liver Imaging Reporting and Data System category 5 for hepatocellular carcinoma: a systematic review and meta-analysis of comparative studies.

Yeun-Yoon KimSunyoung LeeJaeseung ShinWon Jeong SonYun Ho RohJeong Ah HwangJi Eun Lee
Published in: European radiology (2022)
• The pooled sensitivity of LR-5 using MRI was higher than that using CT (61% versus 48%), but the pooled specificities of LR-5 were not significantly different between CT and MRI (96% versus 93%). • Subgroup analysis according to the MRI contrast media showed a significantly higher pooled per-observation sensitivity using ECA-enhanced MRI than with EOB-enhanced MRI (73% versus 55%), and comparable specificities (93% versus 94%). • Although LI-RADS provides a common diagnostic algorithm for CT or MRI, the per-observation performance of LR-5 can be affected by the imaging modality as well as the MRI contrast agent.