Gas Pulse-X-Ray Probe Ambient Pressure Photoelectron Spectroscopy with Submillisecond Time Resolution.

Andrey ShavorskiyGiulio D'AcuntoVirgínia BoixMattia ScardamagliaSuyun ZhuRobert H TempertonJoachim SchnadtJan Knudsen
Published in: ACS applied materials & interfaces (2021)
A setup capable of conducting gas pulse-X-ray probe ambient pressure photoelectron spectroscopy with high time resolution is presented. The setup makes use of a fast valve that creates gas pulses with an internal pressure in the mbar range and a rising edge of few hundreds of microseconds. A gated detector based on a fast camera is synchronized with the valve operation to measure X-ray photoemission spectra with up to 20 μs time resolution. The setup is characterized in several experiments in which the N2 gas is pulsed either into vacuum or a constant flow of another gas. The observed width of the pulse rising edge is 80 μs, and the maximum internal pulse pressure is ∼1 mbar. The CO oxidation reaction over Pt (111) was used to demonstrate the capability of the setup to correlate the gas phase composition with that of the surface during transient supply of CO gas into an O2 stream. Thus, formation of both chemisorbed and oxide oxygen species was observed prior to CO gas perturbation. Also, the data indicated that both the Langmuir-Hinshelwood and Mars-van-Krevelen mechanisms play an important role in the oxidation of carbon monoxide under ambient conditions.