New generation cardioverter-defibrillator lead with a floating atrial sensing dipole: Long-term performance.

Erdal SafakGiuseppe D'AnconaHilmi KaplanEvren CaglayanStephan KischeAlper ÖnerHüseyin InceJasmin Ortak
Published in: Pacing and clinical electrophysiology : PACE (2018)
Although the Linox S DX lead presents a satisfactory long-term stability of the atrial sensing, many of the messages sent to the device during follow-up were incorrectly classified by the ICD software. As a result, inappropriate therapy occurred with a rate similar to that observed with dual-chamber ICDs, but some malignant arrhythmias remained undiagnosed and untreated while occurring. Results in larger prospective cohorts should be analyzed and software improvements of the device should be suggested to overcome these potential drawbacks.