Improving puncture accuracy in percutaneous CT-guided needle insertion with wireless inertial measurement unit: a phantom study.

Chia-Ying LinWen-Ruei TangPo-Chang ChiangJenn-Jier James LienPei-Yi TsengYi-Sheng LiuChao-Chun ChangYau-Lin Tseng
Published in: European radiology (2023)
• The IMU-assisted method showed a significant decrease in the number of needle passes (IMU 1.2 ± 0.42, laser protractor 2.9 ± 1.6, freehand 3.6 ± 2.0 time, p < 0.001). • The IMU-assisted method showed a significant decrease in the procedure time (IMU 3.0 ± 1.2, laser protractor 6.4 ± 2.9, freehand 6.2 ± 3.1 min, p < 0.001). • The IMU-assisted method showed a significant decrease in the mediolateral angle error of the first pass and the caudocranial angle error of the first pass.