The novel MR-DN system was safe and time-efficient, as the surgery took 30 min from anaesthesia to suturing. The accuracy of implant placement was high with minimal deviations recorded in the three planes of space compared to the presurgical planning: the error at the entry point planar distance (XY) was 0.381 mm, and the entry point planar distance (Z) was 0.173 mm, for a 3D entry point distance (En) of 0.417 mm. A 3D apex deviation (An) of 0.193 mm was registered, with an angle difference of 1.852° Conclusions: This proof-of-concept study demonstrated the clinical feasibility of MR-DN for guided implant placement in single tooth gaps. Further clinical studies on a large sample of patients are needed to confirm these positive preliminary results. Statement of clinical relevance: The use of MR-DN can change the perspectives of guided dental implant surgery as a possible alternative to the classic static and dynamic guided surgical techniques for the rehabilitation of single tooth gaps.