Extending coherence time of macro-scale diamond magnetometer by dynamical decoupling with coplanar waveguide resonator.
Yuta MasuyamaK MizunoH OzawaH IshiwataY HatanoTakeshi OhshimaT IwasakiM HatanoPublished in: The Review of scientific instruments (2019)
Ultimate sensitivity for quantum magnetometry using nitrogen-vacancy (NV) centers in a diamond is limited by a number of NV centers and coherence time. Microwave irradiation with a high and homogeneous power density for a large detection volume is necessary to achieve a highly sensitive magnetometer. Here, we demonstrate a microwave resonator to enhance the power density of the microwave field and an optical system with a detection volume of 1.4 × 10-3 mm3. The strong microwave field enables us to achieve 48 ns Rabi oscillation which is sufficiently faster than the phase relaxation time of NV centers. This system combined with a decoupling pulse sequence, XY16, extends the spin coherence time (T 2) up to 27 times longer than that with a spin echo method. Consequently, we obtained an AC magnetic field sensitivity of 10.8 pt/ Hz using the dynamical decoupling pulse sequence.