Login / Signup

Effect of Gaussian Smoothing Filter Size for CT-Based Attenuation Correction on Quantitative Assessment of Bone SPECT/CT: A Phantom Study.

Yoya TomitaYasutaka IchikawaKengo HashizumeHajime Sakuma
Published in: Journal of digital imaging (2023)
This study aims to determine the effect of Gaussian filter size for CT-based attenuation correction (CTAC) on the quantitative assessment of bone SPECT. An experiment was performed using a cylindrical phantom containing six rods, of which one was filled with water and five were filled with various concentrations of K 2 HPO 4 solution (120-960 mg/cm 3 ) to simulate different bone densities. 99m Tc-solution of 207 kBq/ml was also included within the rods. SPECT data were acquired at 120 views for 30 s/view. CT for attenuation correction were obtained at 120 kVp and 100 mA. Sixteen different CTAC maps processed with different Gaussian filter sizes (ranging from 0 to 30 mm in 2 mm increments) were generated. SPECT images were reconstructed for each of the 16 CTAC maps. Attenuation coefficients and radioactivity concentrations in the rods were compared with those in the water-filled rod without K 2 HPO 4 solution as a reference. Gaussian filter sizes below 14-16 mm resulted in an overestimation of radioactivity concentrations for rods with high concentrations of K 2 HPO 4 (≥ 666 mg/cm 3 ). The overestimation of radioactivity concentration measurement was 3.8% and 5.5% for 666 mg/cm 3 and 960 mg/cm 3 K 2 HPO 4 solutions, respectively. The difference in radioactivity concentration between the water rod and the K 2 HPO 4 rods was minimal at 18-22 mm. The use of Gaussian filter sizes smaller than 14-16 mm caused an overestimation of radioactivity concentration in regions of high CT values. Setting the Gaussian filter size to 18-22 mm enables radioactivity concentration to be measured with the least influence on bone density.
Keyphrases
  • image quality
  • dual energy
  • computed tomography
  • contrast enhanced
  • bone mineral density
  • pet ct
  • positron emission tomography
  • soft tissue
  • bone loss
  • bone regeneration
  • magnetic resonance imaging
  • magnetic resonance