Colorimetric and Reverse Fluorescence Dual-Signal Readout Immunochromatographic Assay for the Sensitive Determination of Sibutramine.
Yun GuiYun ZhaoPengyan LiuYulong WangXinxin MaoChi-Fang PengBruce D HammockCunzheng ZhangPublished in: ACS omega (2024)
Later flow immunochromatographic assay has been widely used in clinical, environmental, and other diagnostic applications owing to its high sensitivity and throughput. However, most immunoassays operate in the "turn-off" mode for detecting targets of low molecular weight. The signal intensity decreases as the analyte concentration increases, which poses a challenge for achieving ultrasensitive detection at low concentrations and is counterintuitive to new users. In this work, a fluorometric immunochromatographic assay (FICA) is developed to simultaneously read "turn-on" fluorescent and "turn-off" colorimetric signals, where ZnCdSe/ZnS quantum dots act as fluorescence donors and gold nanoparticles (AuNPs) act as quenchers. The fluorescent signal (excitation/emission wavelengths of 365/525 nm) is positively correlated with analytes' concentration. Taking sibutramine (SBT) as the analysis target, the visual limit of detection for SBT reached 3.9 ng/mL, and the limit of Quantitation was 5.0 ng/mg in spiked samples. The developed FICA achieves a high sensitivity in SBT detection, which is much lower than that of the colloidal gold-based immunochromatographic assay. This dual-function detection mode has great potential to be used as a rapid on-site semiquantitative method, providing an alternative mode for the determination of low levels of target analytes.
Keyphrases
- quantum dots
- sensitive detection
- loop mediated isothermal amplification
- gold nanoparticles
- label free
- living cells
- fluorescent probe
- energy transfer
- high throughput
- single molecule
- molecularly imprinted
- real time pcr
- solid phase extraction
- mass spectrometry
- photodynamic therapy
- risk assessment
- high intensity
- kidney transplantation
- liquid chromatography