Less early subsidence of cemented Exeter short stems compared with cemented Exeter standard stems in Dorr type A femurs.
Peter Bo JørgensenStig Storgaard JakobsenDovydas VainoriusMorten HomiliusTorben Bæk HansenMaiken StillingPublished in: Bone & joint open (2023)
The standard Exeter stem had more early subsidence compared with the short Exeter stem in patients with Dorr type A femora, but thereafter a similar migration pattern of subsidence until minimum five years follow-up. Both the standard and the short Exeter stems subside. The standard stem subsides more compared to the short stem in Dorr type A femurs. Subsidence of the Exeter stems was not affected by cement mantle thickness.