Analytical Performance of the New Siemens Affinity Chrome-Mediated Immunoassay Everolimus Assay and Its Interchangeability With the Thermo Quantitative Microsphere System for Routine Therapeutic Drug Monitoring of Patients After Solid Organ Transplantation.
Cristiano IalongoMaria SapioAntonio AngeloniPublished in: Therapeutic drug monitoring (2023)
ACMIA/EVRO shows satisfactory analytical performances that comply with recommendations, but it does not fulfill requirements for interchangeability with QMS/EVER. Particularly, this new assay using sirolimus-specific antibody shows a sizable proportional bias versus the more specific comparator, which may be because of EVL metabolites. This is supported by the lack of agreement for individual differences in most samples collected at the peak concentration (C2). Therefore, further evidence is needed to support the transition of EVL level monitoring from QMS/EVER to ACMIA/EVRO without making extensive changes to both reference interval and patient's baseline.