Doppler Velocity Estimates of Internal Carotid Artery Stenosis: Angle Correction Parallel to the Color Doppler Lumen Versus Parallel to the Artery Wall.
Joseph F PolakJean M Alessi-ChinettiFrederick W KremkauPublished in: Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine (2019)
Blood flow velocity measurements made with angle correction parallel to the direction of the color Doppler blood flow lumen remain more consistent with the distance from the stenosis than blood flow velocity measurements made with angle correction parallel to the outer artery walls.