In Compton PET, that has a scatterer inserted inside a PET ring, there are multi-interaction events that can be treated as both PET and Compton events. A PET event from multi-interaction events that include a Compton event and a photoelectric absorption event or two Compton events can be extracted by applying a PET recovery method. In this study, we aimed to establish a method to maximize image quality by utilizing such redundant events. We conducted brain-scale Monte Carlo simulations of a C-shaped Compton-PET geometry and a whole gamma imaging (WGI) geometry. Images were reconstructed by a hybrid image reconstruction method combining both PET and Compton events. The result showed that the spatial resolution was improved when treated as PET events while keeping the noise level. The effect of improvement was more significant in WGI than in C-shaped Compton PET because the number of events recovered as PET events having more accurate spatial information was much larger in WGI. When the PET-recovered multi-interaction events were also included as Compton events in the hybrid reconstruction, we did not observe any improvement in image quality, while the number of used events was largest. The results suggested that treating events as PET events exclusively was better for image quality.