Direct comparison of Xpert Xpress, FilmArray Respiratory Panel, Lumipulse antigen test, and RT-qPCR in 165 nasopharyngeal swabs.
Yosuke HirotsuMakoto MaejimaMasahiro ShibusawaYume NatoriYuki NagakuboKazuhiro HosakaHitomi SuekiKenji AmemiyaMiyoko HayakawaHitoshi MochizukiToshiharu TsutsuiYumiko KakizakiYoshihiro MiyashitaMasao OmataPublished in: BMC infectious diseases (2022)
Xpert Xpress and FilmArray Respiratory Panel exhibited an equivalent performance. The Lumipulse antigen test was slightly less sensitive than the NAATs, but showed high assay performance except for samples with low viral load. The Xpert Xpress, FilmArray Respiratory Panel and Lumipulse antigen tests offer rapid sample-to-answer data, allowing random access detection on automated devices.