Prospective randomized clinical trial evaluating the effects of two different implant collar designs on peri-implant healing and functional osseointegration after 25 years.
Aldo Joseph CamardaRobert DurandMarwa BenkarimPierre H RompréGeneviève GuertinHugo CiaburroPublished in: Clinical oral implants research (2021)
Implants with the 1.2mm machined collar limited bone loss to 1mm, while those with the longer machined collar showed > 1.5mm loss after 25 years of function with microgap at the crest. Internal-connection design and fixture surface microtexturing did not result in greater bone preservation. ClinicalTrials.gov Identifier: NCT03862482.