Finite element analysis of 3D-printed personalized titanium plates for mandibular angle fracture.

Yan LiQingguo LaiQingguo LaiRunqi XueKaiwen ZhuYanwei Deng
Published in: Computer methods in biomechanics and biomedical engineering (2022)
This paper discussed the size of 3 D-printed personalized titanium plates that can gain maximum stability of mandibular fracture and minimize stress shielding through finite element analysis. A 3 D virtual model of mandible with mandibular angle fracture was created from the CT data of patient. 3 D-printed personalized titanium plates varying in length and thickness were designed, and finite element analysis was performed under different loading conditions and fracture healing periods. After that, the stress distribution and deformation of the mandible with gonial fracture could be observed, and the stress shielding rate could be obtained. Finally, SPSS21.0 was used for statistical analysis. The results of finite element analysis indicated that as the thickness of titanium plates and the healing time decreased, the maximum displacement increased, under a certain condition in which the pore size, the width, the hole distance and the bridge spacing were 2.0 mm, 4.0 mm, 6.0 mm, 12.0 mm, respectively. What's more, in this condition, the thicker the titanium plate and the shorter the healing time were, the higher the stress shielding was at central occlusion. When the thickness of the personalized 3 D-printed titanium plate was 1.0 mm, the maximum displacement tended to be stable and the stress shielding was minimized. It can not only improve the bone stability after tension band fixation, but also minimize the stress shielding, which is expected to expand the indications of tension band fixation.
Keyphrases
  • finite element analysis
  • stress induced
  • minimally invasive
  • hip fracture
  • computed tomography
  • magnetic resonance imaging
  • heat stress