Efficient Nanozyme-Triggered Pressure Sensor for Point-of-Care Immunoassay: Visual Sensing and Time Readout Device.

Mengting ChenQiqian QiuAori QilengHaoran ShenWeipeng LiuYingju Liu
Published in: Analytical chemistry (2023)
Point-of-care testing (POCT), with its portability and high sensitivity, is an analytical device for rapid on-site sensing and detection. In this study, a POCT device was designed for the portable detection of illegal additives by integrating a coil device that can visually sense color distance and a two-electrode electrochemical system. Real-time monitoring of pressure changes was achieved by driving CeO 2 @Pt/Au nanoparticle (NP)-labeled antibodies into a competitive immunoreaction, in which CeO 2 and Pt/Au synergistically catalyzed the production of large amounts of O 2 from H 2 O 2 , leading to a significant increase in gas within the closed chamber. Attractively, the coil device converted the pressure stimulus into visually readable change in distance for semi-quantitative detection of the target substance, while the electrical signal output caused by the changes of the solution around the electrodes achieved accurate and reliable quantification of the target. In addition, the proposed dual-mode pressure immunoassay device has acceptable selectivity, stability, and reproducibility. Herein, this portable device, which enables target concentration readings by converting pressure into multiple signals, provides an effective way to visualize POCT assays in resource-limited areas.