Login / Signup

β-Hydroxybutyrate dehydrogenase decorated MXene nanosheets for the amperometric determination of β-hydroxybutyrate.

Aneesh KoyappayilSachin Ganpat ChavanMohsen MohammadniaeiAnna GoSei Young HwangTaesung Kim
Published in: Mikrochimica acta (2020)
MXene nanosheets of type Ti3C2Tx were modified with β-hydroxybutyrate dehydrogenase and then used as a biosensor for amperometric sensing of β-hydroxybutyrate. The MXene and the nanocomposite were characterized by X-ray photoelectron spectroscopy, field-emission scanning electron microscopy, X-ray diffraction, and Fourier transform infrared spectroscopy. The MXene has a layered structure and proved to be an excellent immobilization matrix providing good compatibility with the enzyme β-hydroxybutyrate dehydrogenase. The MXene-based biosensor, best operated at a potential of - 0.35 V (vs. Ag/AgCl), displays a wide linear range (0.36 to 17.9 mM), a sensitivity of 0.480 μA mM-1 cm-2, and a low detection limit (45 μM). The biosensor was successfully applied to the determination of β-hydroxybutyrate in (spiked) real serum samples. Graphical abstract Schematic representation of the synthesis and decoration of Mxene 2D sheets with β-hydroxybutyrate dehydrogenase for the amperometric determination of β-hydroxybutyric acid.
Keyphrases