Tissue oxygenation indices of cerebrovascular autoregulation in healthy volunteers: a comparison of two NIRS devices.
Alexander PengMatthew J ChuaBernard ChanMatthias JaegerAnders ÅnemanAlwin ChuanPublished in: Neurological research (2020)
The substantial bias and wide limits of agreement for the tissue oximetry index as a measure of cerebral autoregulation indicate that values must be treated as specific to models of near-infrared spectrometers.