Optimal relaxation parameters of dynamic row-action maximum likelihood algorithm and post-smoothing filter for image reconstruction of dedicated breast PET.
Yoko SatohMasamichi ImaiKenji HirataYuzo AsakawaChihiro IkegawaHiroshi OnishiPublished in: Annals of nuclear medicine (2021)
The relaxation coefficient factor β20 and a 0.78- or 1.17-mm post-filter were optimal for dbPET image reconstruction with balanced spatial resolution and noise. However, they should be selected according to the impact on the dbPET image and the purpose of the examination.