Escape of surgical smoke particles, comparing conventional and valveless trocar systems.
Daniel RobertsonFrank SterkeWillem van WeteringenAlberto ArezzoYoav Mintznull nullTim HoremanPublished in: Surgical endoscopy (2023)
This study shows that a valveless trocar system releases more particles into the operating room environment than a conventional trocar. During instrument insertion, the leakage through the valveless system is 6 to 20 times higher than the conventional system. With a valveless trocar, leakage decreases with increasing pressure. With both trocar types leakage depends on instrument diameter.