RF-induced heating for active implantable medical devices in dual parallel leads configurations at 1.5 T MRI.
Wei HuRan GuoQingyan WangJianfeng ZhengJeffrey TsangWolfgang KainzStuart LongJi ChenPublished in: Magnetic resonance in medicine (2023)
For the AIMD system in this study, it shows that the coupling effects between the dual parallel leads of AIMD systems can reduce RF-induced heating. The global transfer function for different spatial distance dual-lead configurations can potentially provide a method for the RF-induced heating evaluation for dual-lead AIMD systems.