Login / Signup

Sensitivity and specificity of thromboelastography for hyperfibrinolysis: Comparison of TEG 5000 and TEG 6S CK LY30 systems.

Christine FujaTimothy Clifford CarllKrzysztof MikrutGeoffrey D Wool
Published in: American journal of clinical pathology (2023)
CK LY30 above the ULN is a sensitive but not specific cutoff for hyperfibrinolysis. At least moderately elevated CK LY30 carries more clinical portent on the TEG 6s instrument than on the TEG 5000. These TEG instruments are not sensitive to low concentrations of tPA.
Keyphrases
  • protein kinase
  • patient reported outcomes