Evaluation of the performance of two new generation pulse oximeters in cats at different probe positions and under the influence of vasoconstriction.
René DoerfeltJulia DielsKatrin HartmannPublished in: Journal of feline medicine and surgery (2021)
The Masimo pulse oximeter had lower signal failure rates at the standard probe position on the tongue and at 2/4 alternative probe positions. The standard probe position on the tongue had the lowest failure rate for both devices. Dexmedetomidine-induced vasoconstriction increased the failure rate for the EDAN but not for the Masimo pulse oximeter.