Reliability of the Parabola Approximation Method in Heart Rate Variability Analysis Using Low-Sampling-Rate Photoplethysmography.
Hyun Jae BaekJaeWook ShinGunwoo JinJaegeol ChoPublished in: Journal of medical systems (2017)
Photoplethysmographic signals are useful for heart rate variability analysis in practical ambulatory applications. While reducing the sampling rate of signals is an important consideration for modern wearable devices that enable 24/7 continuous monitoring, there have not been many studies that have investigated how to compensate the low timing resolution of low-sampling-rate signals for accurate heart rate variability analysis. In this study, we utilized the parabola approximation method and measured it against the conventional cubic spline interpolation method for the time, frequency, and nonlinear domain variables of heart rate variability. For each parameter, the intra-class correlation, standard error of measurement, Bland-Altman 95% limits of agreement and root mean squared relative error were presented. Also, elapsed time taken to compute each interpolation algorithm was investigated. The results indicated that parabola approximation is a simple, fast, and accurate algorithm-based method for compensating the low timing resolution of pulse beat intervals. In addition, the method showed comparable performance with the conventional cubic spline interpolation method. Even though the absolute value of the heart rate variability variables calculated using a signal sampled at 20 Hz were not exactly matched with those calculated using a reference signal sampled at 250 Hz, the parabola approximation method remains a good interpolation method for assessing trends in HRV measurements for low-power wearable applications.