Indirect measurement of absolute cardiac output during exercise in simulated altered gravity is highly dependent on the method.
Richard S WhittleLindsay M StapletonLonnie Grove PetersenAna Diaz-ArtilesPublished in: Journal of clinical monitoring and computing (2021)
There is a poor agreement in absolute stroke volume and cardiac output values between measurement via PCA (Nexfin) and IGR (Innocor) in subjects who are exercising in simulated altered gravity environments. These results suggest that the chosen measurement method and device greatly impacts absolute measurements of cardiac output. However, there is a good level of agreement between the two devices when measuring relative changes. Either of these devices seem adequate to capture cardiac changes, but should not be solely relied upon for accurate measurement of absolute cardiac output.