Mobile 3D C-arm fluoroscopy can be safely incorporated into DBS workflows for both registration and lead verification. For registration, the limited field of view requires the use of frameless transient fiducials and is highly accurate. For lead position verification based on MRI co-registration, we estimate there is around a 0.4 mm discrepancy between lead position seen on 3DXT versus CT when corrected for brain shift. This is similar to that described in O-arm® or mobile CT series. For units where logistical or financial considerations preclude the acquisition of a cone beam CT or mobile CT scanner, our data support portable 3D C-arm fluoroscopy as an acceptable alternative with significantly lower radiation exposure.