Clinical laboratory verification of thyroglobulin concentrations in the presence of autoantibodies to thyroglobulin: comparison of EIA, radioimmunoassay and LC MS/MS measurements in an Urban Hospital.

Sarah E WheelerLi LiuHarry C BlairRichard SivakNancy LongoJeffery TischlerKathryn MulveyOctavia M Peck Palmer
Published in: BMC research notes (2017)
We measured TgAb and Tg in 50 plasma samples from 40 patients in whom Tg measurement was part of their routine follow-up and 10 healthy volunteers. Discrepancy between EIA and both LC-MS/MS and RIA was observed at low Tg concentrations (≤ 7.55 ng/mL) in TgAb positive specimens (LC-MS/MS = 1.9 * EIA - 0.03, r = 0.68). RIA and LC-MS/MS Tg measurements in TgAb positive specimens with low Tg concentrations had improved correlation but demonstrated bias (LC MS/MS = 0.6 * RIA - 1.4, r = 0.90). Disagreement between methods may be attributed to LC-MS/MS reported Tg concentrations as undetectable compared to RIA. It seems likely that most discrepant cases are falsely elevated in RIA due to TgAb interference, however, some cases appear below the detection limit of LC-MS/MS; implementation of LC-MS/MS by clinicians will require lower detection limits.