Mechanical wear analysis helps understand a mechanism of failure in retrieved magnetically controlled growing rods: a retrieval study.
Jack Zijian WeiHarry S HothiHolly MorgantiSean BergiersElisabetta Dal GalDoris LikcaniJohann HenckelAlister J HartPublished in: BMC musculoskeletal disorders (2020)
We found an association between damage to the internal mechanism of the rods and (1) patterns of single-side longitudinal wear marks and (2) increased material loss. As the material loss was also found to increase over time of rod in situ, we emphasise the importance of early detection and revision of failed MAGEC rods in clinical practice.