Login / Signup

Immunosuppressive drugs in whole blood: validation of a commercially available liquid chromatography/tandem mass spectrometry kit and comparison with immunochemical assays.

Elisa PolledriRosa MercadanteChiara Ferraris FusariniRita MaiavaccaSilvia Fustinoni
Published in: Rapid communications in mass spectrometry : RCM (2018)
The LC/MS/MS certified kit was validated for the detection of immunosuppressant drugs in whole blood and it provided a high-throughput method that is consistent with the requirements of clinical laboratories. The comparison of patient data between LC/MS/MS and up-dated immunoassays shows that a significant discrepancy still exists, especially for CSA and SIRO, confirming the greater specificity associated with use of the LC/MS/MS assay Copyright © 2017 John Wiley & Sons, Ltd.
Keyphrases