Erythrocyte sedimentation rate measurements using MIX-RATE® X20 and VISION A automated analyzers: Method validation and comparison study.

Eakachai PrompetcharaSunudda NowaratsoponSuppakorn WongkamchaiJittra SrieakpanitChutitorn Ketloy
Published in: International journal of laboratory hematology (2022)
Overall, results from both automated ESR analyzers showed comparable analytical performance with the Westergren method especially for low ESR levels. However, both positive and negative systematic bias were documented in the high levels. Thus, for clinical use, it must be assessed whether these biases could affect the cut-off for significant clinical values.