Login / Signup

Incorporating plan complexity into the statistical process control of volumetric modulated arc therapy pre-treatment verifications.

Serenella RussoJordi SaezMarco EspositoAndrea BruschiAlessandro GhirelliSilvia PiniSilvia ScocciantiVictor Hernandez
Published in: Medical physics (2024)
Incorporating plan complexity into SPC for VMAT verifications requires meticulous and comprehensive analysis. To ensure overall process control, we advocate for stringent control and minimization of plan complexity during treatment planning, especially when control limits are adjusted based on plan complexity.