Timing, Energy, and Spatial Characterization of Highly Sampled Monolithic PET Detectors with Different Thicknesses.
Celia ValladaresJohn BarrioMarta FreireNeus CucarellaEfthymios LamprouRobert S MiyaokaWilliam C J HunterRobert HarrisonAntonio J GonzalezPublished in: IEEE transactions on radiation and plasma medical sciences (2023)
The HyPET project proposes a hybrid dedicated TOF-PET for prostate imaging, with pixelated detector blocks in the front layer and monolithic blocks in the back layer. In this work, four detector configurations have been experimentally evaluated for the rear detector layer. The detector configuration consists of LYSO monolithic blocks with the same size (25.4 mm × 25.4 mm) but different thicknesses (5, 7.5, 10, and 15 mm) coupled to the same SiPM array. Each detector configuration has been experimentally characterized in terms of time, energy and spatial resolution by scanning the crystal surface using a fan beam in steps of 0.25 mm. Regarding spatial resolution, the interaction position was estimated using a Neural Network technique. All resolutions except energy, which remains nearly constant at 17% for all cases, show better values for the 5 mm detector thickness. We have achieved spatial resolution values of FWHM of 1.02 ± 0.10, 1.19 ± 0.13, 1.53 ± 0.17, 2.33 ± 0.55 mm, for the 5, 7.5, 10, and 15 mm blocks, respectively. The detector time resolution obtained was 275 ± 26, 291 ± 21, 344 ± 48, and 433 ± 45 ps respectively, using the energy weighted average method for the time stamps.