Application of a new volumetric microsampling device for quantitative bioanalysis of immunosuppression.

Abbie D LeinoJohn Takyi-WilliamsBo WenDuxin SunManjunath Amit P Pai
Published in: Bioanalysis (2022)
Background: Volumetric absorptive microsampling may reduce the blood collection burden associated with therapeutic drug monitoring of immunosuppression to prevent organ transplant rejection. This work describes the development of a laboratory and analytical technique for quantifying tacrolimus and mycophenolic acid (MPA) from the Tasso-M20™ in human whole blood using bead-based impact-assisted extraction. Results: The sampled blood volume was accurate with estimated volumes within <2% of the expected 20 μl. Recovery using impact-assisted extraction was 73-87% for MPA and 100% for tacrolimus and was hematocrit-independent for both analytes. The LC-MS/MS assay is precise and accurate within the acceptance criteria of 15%. Conclusion: The sampling and extraction procedures allowed for accurate quantification of tacrolimus and MPA. Exploration of abuse scenarios identified important education points for patients conducting home-based sample collections in the future.