Towards a practical methodology for assessment of the objective occlusion effect induced by earplugs.
Hugo Saint-GaudensHugues NélisseFranck SgardOlivier DoutresPublished in: The Journal of the Acoustical Society of America (2022)
The occlusion effect (OE) occurs when the earcanal becomes occluded by an in-ear device, sometimes leading to discomforts experienced by the users due to the augmented perception of physiological noises, or to a distorted perception of one's own voice. The OE can be assessed objectively by measuring the amplification of the low-frequency sound pressure level (SPL) in the earcanal using in-ear microphones. However, as revealed by methodological discrepancies found in past studies, the measurement of this objective occlusion effect (OE obj ) is not standardized. With the goal of proposing a robust yet simple methodology adapted for field assessment, three experimental aspects are investigated: (i) stimulation source and the stimulus's characteristics to induce the phenomenon, (ii) measurement method of the SPL in earcanal, (iii) indicator to quantify the OE obj . To do so, OE obj is measured on human participants in laboratory conditions. Results obtained with a specific insert device suggest using the participant's own voice combined with simultaneous measurements of the SPLs based on the noise reduction method and using a single value indicator leads to a simple yet robust methodology to assess OE obj . Further research is necessary to validate the results with other devices and to generalize the methodology for field assessment.