Reliability of a point of care testing blood gas analyzer for measurement of lactate levels in cerebrospinal fluid.

Alberto Fernández ReinaMaría Dolores López AbellánSamir Attaibi HadriÁngela Puche CandelMaría Isabel Díaz LópezEva Pérez FernándezManuel Tomás Orgaz MoralesLuis García de Guadiana Romualdo
Published in: Scandinavian journal of clinical and laboratory investigation (2024)
Analysis of cerebrospinal fluid (CSF), including lactate, is key for diagnosis of acute meningitis. Since blood gas analyzers (BGA) enable rapid and safe blood-lactate measurements, we evaluated the reliability of RAPIDPoint 500 BGA to provide a fast and accurate measure of CSF lactate. In this study, CSF lactate levels were measured by a reference assay and on RAPIDPoint 500 BGA. Comparability was evaluated through difference analysis, using Bland Altman test, and linear regression analysis, using the Passing Bablok test. Agreement rate according to CSF lactate (≥3.5 and <3.5 mmol/L) was calculated using kappa (κ) statistic. Population study included 98 CSF samples. Concerning difference analysis, according to Bland-Altman test, bias was 0.13 mmol/L (CI 95%: -0.26 to 0.52 mmol/L. In regression analysis, according to Passing-Bablok equation a systematic difference between both assays was found. In concordance analysis, the interrate realibility was very high (κ: 0.964). According to our resuls, although a systematic difference was detected when lactate levels were measured on RAPIDPoint 500 BGA, the results from Bland-Altman test and the high agreement rate support that this POCT analyzer could be useful for a early and safe detection of patients with high probability of increased CSF lactate level.
Keyphrases
  • cerebrospinal fluid
  • loop mediated isothermal amplification
  • room temperature
  • nuclear factor
  • inflammatory response
  • acute respiratory distress syndrome
  • ionic liquid