A multichannel gated neutron detector with reduced afterpulse for low-yield neutron measurements in intense hard X-ray backgrounds.

Yuki AbeN NakajimaY SakaguchiY ArikawaS R MirfayziShinsuke FujiokaT TaguchiK MimaA YogoH NishimuraH ShiragaM Nakai
Published in: The Review of scientific instruments (2018)
A design of multichannel gated photomultiplier tube (PMT) is presented for the 960-channel neutron time-of-flight detector at the Institute of Laser Engineering of Osaka University. This is important for the fusion science and the nuclear photonics where intense hard X-rays are generated from the interaction of ultra-short laser pulse of petawatt power density with matter. The hard X-rays often overload PMTs and cause signal-induced background noises called afterpulses, making the detection of subsequent neutrons impossible. For this reason, the PMTs are coupled with an electrical time-gating (ETG) system to avoid overloading. The ETG system disables the PMT by modulating the dynode potential during the primary X-ray flash. An after-pulsing suppression technique is demonstrated by applying a reverse bias voltage between the photocathode and the first dynode. The presented multichannel scheme provides a gate response time of 80 ns, a signal cutoff ratio of 2.5 × 102, and requires reasonably low power consumption.