Login / Signup

[Effects of automated external defibrillators on hands-off intervals in lay rescuers].

Volker SchäferPatrick WitwerLisa SchwingshacklHannah SalchnerLukas GasteigerWilfried SchabauerWolfgang Lederer
Published in: Notfall & rettungsmedizin (2022)
Perishock pause of less than 10 s was not achieved with any of the tested devices. Shortened and more precise voice prompts as well as more clearly arranged labeling and layout of pads are needed to simplify application, reduce delayed onset of chest compressions and shorten hands-off intervals.
Keyphrases
  • machine learning
  • deep learning
  • high throughput
  • cardiopulmonary resuscitation
  • cardiac resynchronization therapy
  • cardiac arrest
  • heart failure