Heart Rate-Dependent Degree of Motion Artifacts in Coronary CT Angiography Acquired by a Novel Purpose-Built Cardiac CT Scanner.

Milán Vecsey-NagyÁdám Levente JermendyMarton KolossvaryBorbála VattayMelinda BoussoussouFerenc Imre SuhaiAlexisz PanajotuJudit CsőreSarolta BorzsákDaniele Mariastefano FontaniniCsaba Csobay-NovákBéla MerkelyPál Maurovich-HorvatBálint Szilveszter
Published in: Journal of clinical medicine (2022)
Although reaching target heart rate (HR) before coronary CT angiography (CCTA) is still of importance, adequate HR control remains a challenge for many patients. Purpose-built cardiac scanners may provide optimal image quality at higher HRs by further improving temporal resolution. We aimed to compare the amount of motion artifacts on CCTA acquired using a dedicated cardiac CT (DCCT) compared to a conventional multidetector CT (MDCT) scanner. We compared 80 DCCT images to 80 MDCT scans matched by sex, age, HR, and coronary dominance. Image quality was graded on a per-patient, per-vessel and per-segment basis. Motion artifacts were assessed using Likert scores (1: non-diagnostic, 2: severe artifacts, 3: mild artifacts, 4: no artifacts). Patients were stratified into four groups according to HR (<60/min, 60-65/min, 66-70/min and >70/min). Overall, 2328 coronary segments were evaluated. DCCT demonstrated superior overall image quality compared to MDCT (3.7 ± 0.4 vs. 3.3 ± 0.7, p < 0.001). DCCT images yielded higher Likert scores in all HR ranges, which was statistically significant in the 60-65/min, 66-70/min and >70/min ranges (3.9 ± 0.2 vs. 3.7 ± 0.2, p = 0.008; 3.5 ± 0.5 vs. 3.1 ± 0.6, p = 0.048 and 3.5 ± 0.4 vs. 2.7 ± 0.7, p < 0.001, respectively). Using a dedicated cardiac scanner results in fewer motion artifacts, which may allow optimal image quality even in cases of high HRs.
Keyphrases