Disintegration of Multiple Parts of a Bone Transport Nail During Planned Removal: A Case Report.
Ryan T HalvorsonUtku KandemirPublished in: JBJS case connector (2024)
When planning for implant removal, surgeons should be aware of potential intraoperative disconnection of subcomponents of this magnetic bone transport nail and ensure that equipment for retrieval (e.g., very long endoscopy forceps) is available.