Physical limits to human brain B0 shimming with spherical harmonics, engineering implications thereof.
Bruno Pinho MenesesAlexis AmadonPublished in: Magma (New York, N.Y.) (2022)
We show "perfect" SH shimming is possible only if the ROI can be contained in a sphere that does not enclose sources of magnetic field inhomogeneity, which are gathered at the air-tissue interface. We establish a [Formula: see text]Hz inhomogeneity hard shim limit at 7T for whole brain SH shimming, that can only be attained at shimming degree higher than 90. On the other hand, under limited power and SH degree resources, 3D region-specific shimming is shown to greatly improve homogeneity in critical zones such as the prefrontal cortex and around ear canals.