An arteriovenous mock circulatory loop and accompanying bond graph model for in vitro study of peripheral intravascular bioartificial organs.
Jarrett C MoyerVenkat Keshav ChivukulaParsa Taheri-TehraniSukhveer SandhuCharles BlahaWilliam H FissellShuvo RoyPublished in: Artificial organs (2023)
The arteriovenous MCL was a cost-effective, portable system that reproduced flow rates and pressures consistent with a pulsatile arteriovenous shunt as measured in swine. MCL performance was stable over prolonged use, providing a cost-effective simulator for enhanced testing of peripherally implanted bioartificial organ prototypes. The corresponding bond graph model recapitulates MCL and animal physiology, offering a tool for further refinement of the MCL system.