Can We Completely Trust in Automated Software for Preoperative Planning of Shoulder Arthroplasty? Software Update May Modify Glenoid Version, Glenoid Inclination and Humeral Head Subluxation Values.

Raffaele GarofaloAlberto FontanarosaAlessandro CastagnaNunzio LassandroAngelo Del BuonoAngelo De Crescenzo
Published in: Journal of clinical medicine (2023)
The present study shows the software update of an automated preoperative planning program may significantly modify the values of glenoid inclination. Even though without a significant difference, variations were also found for the glenoid version and humeral head subluxation. Accordingly, these results should further advise surgeons to carefully and critically evaluate data acquired with automated software.