Effect of material type, torque value, and sterilization on linear displacements of a scan body: An in vitro study.

Emre DikerHakan TerziogluDiogo N M GouveiaMustafa Borga DönmezJeremy SeidtBurak Yilmaz
Published in: Clinical implant dentistry and related research (2023)
Axial displacement of SBs was affected by material type, torque value, and sterilization. Ti SBs mostly had lower displacements than PEEK SBs. Application of 15 Ncm torque to tested PEEK SBs should be refrained from and a calibrated tightening tool may enable the application of 10 Ncm or lower torque values for lower displacements. Sterilization generally increased PEEK SB displacements.